  Datasheet File OCR Text:
  for raman spectrophotometry, high resolution type (spectral resolution: 0.3 nm) c11713ca c11714cb tg series mini-spectrometers www.hamamatsu.com 1 the c11713ca and c11714cb are polychromators integrated with optical elements, an image sensor and a driver circuit. light to be measured is guided into the entrance port of tg series through an optical ber and the spectrum measured with the built-in image sensor is output from the usb port to a pc for data acquisition. the c11713ca has sensitivity in a wavelength range of 500 to 600 nm, while the c11714cb covers a range of 790 to 920 nm. both types offer a spectral resolution of 0.3 nm. their products come supplied with free evaluation software that allows setting measurement conditions, acquiring and saving data, and displaying graphs. original measurement software can be designed on an end-user?s side as dll?s function speci cation is disclosed. raman spectrophotometry applications * 1: a conversion factor for converting the image sensor pixel number into a wavelength is recorded in the module. a calculation factor for converting the a/d converted count into the input light intensity is not provided. high throughput due to transmission grating made of quartz high accuracy optical characteristics: spectral resolution 0.3 nm features wavelength conversion factor is recorded in internal memory * 1 easy to install into equipment due to compact design selection guide (typ.) parameter c11713ca c11714cb unit spectral response range 500 to 600 790 to 920 nm spectral resolutiion 0.3 nm built-in image sensor back-thinned ccd image sensor ir-enhanced back-thinned ccd image sensor - using a ccd image sensor with enhanced near infrared sensitivity, the c11714cb offers higher sensitivity than that of the previous product. high sensitivity in near infrared region (c11714cb) spectral response kaccb0371ea 750 800 850 900 950 0 100 90 80 70 60 50 40 30 20 10 wavelength (nm) relative photosensitivity (%) c11714cb previous product (typ. ta=25 c)
mini-spectrometers c11713ca, c11714cb tg series 2 electrical and optical characteristics (ta=25 c, unless otherwise noted) parameter c11713ca c11714cb unit min. typ. max. min. typ. max. spectral response range 500 to 600 790 to 920 nm spectral resolution (fwhm) * 6 - 0.3 0.5 - 0.3 0.5 nm wavelength reproducibility * 7 -0.1 to +0.1 -0.1 to +0.1 nm wavelength temperature dependence -0.04 to +0.04 -0.04 to +0.04 nm/c spectral stray light * 6 * 8 - - -30 - - -30 db * 6: depends on the slit opening. values were measured with the slit listed in the table ? structure?. *7: measured under constant light input conditions *8: when monochromatic light of the following wavelengths is input, spectral stray light is de ned as the ratio of the count measured at the input wavelength, to the count measured in a region of the input wavelength 10 nm c11713ca: 550 nm, c11714cb: 860 nm electrical characteristics (ta=25 c, unless otherwise noted) parameter min. typ. max. unit integration time 10 to 10000 ms consumption current of usb bus power * 5 - - 150 ma external power supply voltage - 5 - v consumption current of external power supply * 5 - - 0.8 a * 5: when operated with the supplied evaluation software (at default settings, dark state, excluding start-up). absolute maximum ratings parameter value unit external power supply voltage 5.25 v operating temperature * 4 +5 to +40 c storage temperature * 4 -20 to +70 c *4: no dew condensation when there is a temperature difference between a product and the surrounding area in high humidity environment, dew condensatio n may occur on the product surface. dew condensation on the product may cause deterioration in characteristics and reliability. note: exceeding the absolute maximum ratings even momentarily may cause a drop in product quality. always be sure to use the product within the absolute maximum ratings. structure parameter c11713ca c11714cb unit dimensions (w d h) 120 70 60 mm weight 592 g image sensor back-thinned type ccd image sensor (s10420-1106-01) ir-enhanced back-thinned ccd image sensor (s11510-1006) - number of pixels 2048 1024 pixels slit * 2 (h v) 10 1000 m na* 3 0.11 - connector for optical ber sma905d - a/d conversion 16 bit interface usb 1.1 external power supply (ac adapter) ac 100v-240v, 50/60hz c * 2: entrance slit aperture size * 3: numeric aperture (solid angle)
mini-spectrometers c11713ca, c11714cb tg series 3 spectral resolution vs. wavelength spectral response ( typical example) wavelength (nm) spectral resolution (nm) 450 550 600 500 650 0.1 0 0.2 0.3 0.4 0.5 (typ. ta=25 c) wavelength (nm) relative sensitivity (%) 450 550 600 500 650 20 10 30 50 70 90 40 60 80 0 100 (typ. ta=25 c) wavelength (nm) relative sensitivity (%) 750 850 900 800 950 20 40 60 80 0 100 (typ. ta=25 c) 10 30 50 70 90 0 wavelength (nm) spectral resolution (nm) 750 950 800 850 900 0 0.5 0.4 0.3 0.2 0.1 (typ. ta=25 c) kaccb0224ea kaccb0240ea kaccb0372ea kaccb0373ea c11713ca c11713ca c11714cb c11714cb
mini-spectrometers c11713ca, c11714cb tg series 4 linearity ( typical example ) dark output vs. ambient temperature ( typical example ) 10 100 1000 10000 a/d output difference between ideal value and typical example (%) 100000 -10 -6 -8 10000 1000 integration time (ms) 100 10 -2 -4 0 4 2 6 8 10 typical example of a/d output ideal value of a/d output difference between ideal value and typical example 0 50 40 30 20 10 20000 10000 0 50000 60000 30000 40000 70000 ambient temperature (c) a/d output (averaged over all pixels) (integration time 1 s) a/d output difference between ideal value and typical example (%) integration time (ms) 10 100 1000 10000 100000 -10 -6 -8 10000 1000 100 10 -2 -4 0 4 2 6 8 10 typical example of a/d output ideal value of a/d output difference between ideal value and typical example ambient temperature (c) a/d output (averaged over all pixels) 0 50 40 30 20 10 20000 10000 0 50000 60000 30000 40000 70000 (integration time 1 s) kaccb0235ea kaccb0237ea kaccb0374ea kaccb0375ea c11713ca c11713ca c11714cb c11714cb a/d output is the output with dark output is subtracted when light is input. the difference between the ideal value and typical example contains a measurement error. the smaller the a/d output, the larger the measurement error. a/d output is the sum of the sensor and circuit offset outputs and the sensor dark output.
mini-spectrometers c11713ca, c11714cb tg series 5 optical component layout tg series mini-spectrometers use a transmission holographic grating made of quartz and precision optical components arranged on a rugged optical base, making it possible to deliver high throughput and highly accurate optical characteristics. input slit focus lens image sensor collimating lens transmission grating kaccc0256ea connection example (transmission light measurement) light to be measured is guided into the entrance port of tg series through an optical fiber and the spectrum measured with the built-in image sensor is output through the usb port to a pc for data acquisition. there are no moving parts inside the unit so stable measurements are obtained at all times. an optical ber that guides light input from external sources allows a exible measurement setup. kaccc0577eb fiber exciting ld light source raman probe measuring cell usb cable ac adapter pc mini- spectrometer
mini-spectrometers c11713ca, c11714cb tg series 6 measurement example 790 800 820 840 860 900 880 920 12000 10000 8000 6000 4000 2000 0 14000 wavelength (nm) a/d counts (pump laser 785 nm/60 mw, connection fiber 200 m core dia., integration time 5 s) kaccb0229ea raman spectrophotometry using nophthalene (c11714cb) evaluation software package (supplied with unit) installing the evaluation software package (spec evaluation.exe) * 8 into your pc allows running the following basic tasks: measurement data acquisition and save measurement condition setup module information acquisition (wavelength conversion factor, polychromator type, etc.) graphic display arithmetic operation pixel number to wavelength conversion comparison calculation with reference data (transmittance, re ectance) dark subtraction gaussian approximation (peak position and count, fwhm) note: two or more mini-spectrometers can be connected and used with one pc simultaneously. the external trigger input function does not work with the evaluation software. if using an external trigger input or designing original application software, the user software must be con gured to support that function. *8: compatible os: microsoft ? windows ? 7 professional sp1 (32-bit, 64-bit) microsoft ? windows ? 8 professional (32-bit, 64-bit) dll for controlling hardware is also provided. you can develop your own measurement programs by using a following software development environment. microsoft ? visual studio ? 2008 (sp1) visual c++ ? microsoft ? visual studio ? 2008 (sp1) visual basic ? note: microsoft, windows, visual studio, visual c++ and visual basic are either registerd trademarks or trademarks of microsoft corporation in the united states and other countries.
mini-spectrometers c11713ca, c11714cb tg series 7 dimensional outline (unit: mm, tolerance unless otherwise noted: 0.5) (4 ) m3.0 tap depth 5 (from backside) 60.0 120.0 70.0 110.0 0.2 50.0 0.2 weight: 592 g kacca0281eb accessories usb cable dedicated software (evaluation software, sample software, dll) ac adapter (for power supply) options (sold separately) coaxial cable for external trigger input a10670 optical bers for light input type no. product name core diameter ( m) speci cation a9763-01 fiber for visible/near infrared range 600 na=0.22, length 1.5 m, connectorized sma905d at both ends dimensional outline (a10670, unit: mm) 26.0 31.7 ?6.4 ?14.5 cable length: 1.5 m cable specification: 1.5qev lemo connector ffa.00.250 (made by lemo) bnc connector bnc-p-1.5 (40) (made by hirose or equivalent) kacca0220eb
mini-spectrometers c11713ca, c11714cb tg series 8 mini-spectrometer/micro-spectrometer lineup c9405cb type no. type spectral resolution max. (nm) spectral resolution max. (nm) spectral response range (nm) c10082ca c10082cah c10082md c10083ca c10083cah c10083md c9404ca c9404cah c11482ga c9913gc c9914gb mini-spectrometer tm series mini-spectrometer tg series mini-spectrometer tg series mini-spectrometer ft series 6 6 8 (=320 to 900 nm) (=320 to 900 nm) 8 3 7 7 8 20 3.5 8 8 9 1* 1* 1* type no. type c11009ma c11010ma for installation into mobile measuring equipment 9 * typ. c11007ma c13053ma c11008ma mini-spectrometer rc series mini-spectrometer rc series c11118ga 200 to 800 320 to 1000 200 to 400 500 to 1100 500 to 1100 900 to 1700 900 to 2550 1100 to 2200 340 to 780 0.4* c13054ma 640 to 1050 340 to 780 640 to 1050 for installation into mobile measuring equipment (ultra-compact) mini-spectrometer ms series micro- spectrometer c11697mb (=550 to 900 nm) 5 c11713ca c11714cb 500 to 600 790 to 920 790 to 920 0.3* 0.3* rc-vis-mos spectrometer head spectrometer head spectrometer head spectrometer head rc-swnir-mos 200 400 600 800 1000 1200 1400 1600 1800 2000 2200 2400 2600 spectral response range (nm) 200 400 600 800 1000 1200 1400 1600 1800 2000 2200 2400 2600 tm-uv/vis-ccd tm-vis/nir-ccd tm-vis/nir-ccd tm-vis/nir-mos tg-uv-ccd tm-uv/vis-ccd tm-uv/vis-mos high sensitivity wide dynamic range high resolution high sensitivity tg-cooled nir- i tg2-nir high sensitivity wide dynamic range high resolution tg-uv-ccd high resolution tg-cooled nir- ii non-cooled type ft-swir-mos- ii compact, thin case low noise (cooled type) low noise (cooled type) rc-vis-mos spectrometer module rc-swnir-mos spectrometer module tg-cooled nir- iii low noise (cooled type) ft2-raman compact, thin case tg-swnir-ccd- ii ir-enhanced tm-vis/nir-mos- ii trigger-compatible tg-raman- i high resolution tg-raman- ii high resolution 8 image sensor image sensor back-thinned ccd image sensor cmos linear image sensor cmos linear image sensor cmos linear image sensor back-thinned ccd image sensor back-thinned ccd image sensor ingaas linear image sensor high-sensitivity cmos linear image sensor cmos linear image sensor ir-enhanced back-thinned ccd image sensor ir-enhanced back-thinned ccd image sensor back-thinned ccd image sensor ir-enhanced cmos linear image sensor ir-enhanced cmos linear image sensor high-sensitivity cmos linear image sensor cmos linear image sensor cmos linear image sensor high-sensitivity cmos linear image sensor type no. type 640 to 1050 c11708ma spectrometer head ms-swnir-mos 20 15 15 spectral response range (nm) 200 400 600 800 1000 1200 1400 1600 1800 2000 2200 2400 2600 spectral resolution max. (nm) image sensor c12666ma c12880ma 340 to 780 340 to 850
cat. no. kacc1225e02 oct. 2015 dn www.hamamatsu.com hamamatsu photonics k.k., solid state division 1126-1 ichino-cho, higashi-ku, hamamatsu city, 435-8558 japan, telephone: (81) 53-434-3311, fax: (81) 53-434-5184 u.s.a.: hamamatsu corporation: 360 foothill road, bridgewater, n.j. 08807, u.s.a., telephone: (1) 908-231-0960, fax: (1) 908-23 1-1218 germany: hamamatsu photonics deutschland gmbh: arzbergerstr. 10, d-82211 herrsching am ammersee, germany, telephone: (49) 8152-375-0, fax: (49) 8152-265-8 france: hamamatsu photonics france s.a.r.l.: 19, rue du saule trapu, parc du moulin de massy, 91882 massy cedex, france, telephone: 33-(1) 69 53 71 00, fax: 33-(1) 69 53 71 10 united kingdom: hamamatsu photonics uk limited: 2 howard court, 10 tewin road, welwyn garden city, hertfordshire al7 1bw, united kingdom, telephone: (44) 1707-294888, fax: (44) 1707-325777 north europe: hamamatsu photonics norden ab: torshamnsgatan 35 16440 kista, sweden, telephone: (46) 8-509-031-00, fax: (46) 8-5 09-031-01 italy: hamamatsu photonics italia s.r.l.: strada della moia, 1 int. 6, 20020 arese (milano), italy, telephone: (39) 02-93581733 , fax: (39) 02-93581741 china: hamamatsu photonics (china) co., ltd.: b1201, jiaming center, no.27 dongsanhuan beilu, chaoyang district, beijing 100020 , china, telephone: (86) 10-6586-6006, fax: (86) 10-6586-2866 product specifications are subject to change without prior notice due to improvements or other reasons. this document has been carefully prepared and the information contained is believed to be accurate. in rare cases, however, there may be inaccuracies such as text errors. before using these products, always contact us for the delivery specification sheet to check the latest specifications. the product warranty is valid for one year after delivery and is limited to product repair or replacement for defects discovered and reported to us within that one year period. however, even if within the warranty period we accept absolutely no liability for any loss caused by natural d isasters or improper product use. copying or reprinting the contents described in this material in whole or in part is prohibited without our prior permission. information described in this material is current as of october, 2015. mini-spectrometers c11713ca, c11714cb tg series 9 related information www.hamamatsu.com/sp/ssd/doc_en.html precautions disclamer mini-spectrometers technical information mini-spectrometers


▲Up To Search▲   

 
Price & Availability of C11714CB-15

All Rights Reserved © IC-ON-LINE 2003 - 2022  

[Add Bookmark] [Contact Us] [Link exchange] [Privacy policy]
Mirror Sites :  [www.datasheet.hk]   [www.maxim4u.com]  [www.ic-on-line.cn] [www.ic-on-line.com] [www.ic-on-line.net] [www.alldatasheet.com.cn] [www.gdcy.com]  [www.gdcy.net]


 . . . . .
  We use cookies to deliver the best possible web experience and assist with our advertising efforts. By continuing to use this site, you consent to the use of cookies. For more information on cookies, please take a look at our Privacy Policy. X